Page 1 - LS 6000 Series

TM247971-FJune 2005Beckman Coulter, Inc.4300 North Harbor Boulevard, Fullerton, CA 92834-3100©Copyright 2005 Beckman Coulter, Inc.Printed in U.S.A.LS

Page 3 - Table of Contents

LS 6000 Scintillation System User’s ManualPN 247971-F 5-22 Data CalculationSingle Photon MonitoringCount Time per Data PointThis feature allows you to

Page 4

LS 6000 Scintillation System User’s ManualPN 247971-F 5-23Data CalculationSingle Photon Monitoring5starts the elapsed time before the sample count is

Page 5

LS 6000 Scintillation System User’s ManualPN 247971-F 5-24 Data CalculationSingle Photon Monitoring

Page 6

LS 6000 Scintillation System User’s ManualPN 247971-F 6-1Isotope/DPM LibrariesThe Isotope Library66Isotope/DPM Libraries6.1 The Isotope LibraryIntrodu

Page 7

LS 6000 Scintillation System User’s ManualPN 247971-F 6-2 Isotope/DPM LibrariesThe Isotope LibraryAccessing the Isotope LibraryTo use the Isotope Libr

Page 8

LS 6000 Scintillation System User’s ManualPN 247971-F 6-3Isotope/DPM LibrariesThe Isotope Library62. Highlight “Print Isotope Library” and press SELEC

Page 9

LS 6000 Scintillation System User’s ManualPN 247971-F 6-4 Isotope/DPM LibrariesThe Isotope LibraryAdding A New Isotope to the Isotope LibraryIsotopes

Page 10

LS 6000 Scintillation System User’s ManualPN 247971-F 6-5Isotope/DPM LibrariesThe Isotope Library61. With the New Isotope Setup Menu displayed and “Sc

Page 11 - Introduction

LS 6000 Scintillation System User’s ManualPN 247971-F 6-6 Isotope/DPM LibrariesThe Isotope LibraryNOTE The windows determined by Automatic Setup are

Page 12 - Operating the Instrument

LS 6000 Scintillation System User’s ManualPN 247971-F 6-7Isotope/DPM LibrariesThe Isotope Library610. With both racks in position in the sample change

Page 13 - Operating Modes

LS 6000 Scintillation System User’s ManualPN 247971-F 1-1IntroductionIntended Use11Introduction1.1 Intended UseThe Beckman LS 6720/6730/6750 Scintilla

Page 14 - 1.3 Specifications

LS 6000 Scintillation System User’s ManualPN 247971-F 6-8 Isotope/DPM LibrariesSetting Up A Quench Curve6.2 Setting Up A Quench CurveThe DPM LibraryTh

Page 15 - Figure 1.1 Standard Vial

LS 6000 Scintillation System User’s ManualPN 247971-F 6-9Isotope/DPM LibrariesSetting Up A Quench Curve6standard set, refer to Counting the Standards

Page 16 - Miniature Vials

LS 6000 Scintillation System User’s ManualPN 247971-F 6-10 Isotope/DPM LibrariesSetting Up A Quench Curve5. If setting up quench curves for applicatio

Page 17 - 1.5 Hazards

LS 6000 Scintillation System User’s ManualPN 247971-F 6-11Isotope/DPM LibrariesSetting Up A Quench Curve68. Load the samples to count into the rear mo

Page 18 - Maintenance

LS 6000 Scintillation System User’s ManualPN 247971-F 6-12 Isotope/DPM LibrariesSetting Up A Quench Curvefrom the set and properly dispose of them. A

Page 19 - Getting Started

LS 6000 Scintillation System User’s ManualPN 247971-F 6-13Isotope/DPM LibrariesSetting Up A Quench Curve61. From the Main Menu, highlight “Isotope/DPM

Page 20

LS 6000 Scintillation System User’s ManualPN 247971-F 6-14 Isotope/DPM LibrariesSetting Up A Quench CurveIf you do not wish to count a background quen

Page 21 - 2.3 Operating Controls

LS 6000 Scintillation System User’s ManualPN 247971-F 6-15Isotope/DPM LibrariesSetting Up A Quench Curve6NOTE A message is displayed if a quench curv

Page 22 - Operator Control Key pad

LS 6000 Scintillation System User’s ManualPN 247971-F 6-16 Isotope/DPM LibrariesSetting Up A Quench Curveentered for each isotope. If the dpm in the s

Page 23

LS 6000 Scintillation System User’s ManualPN 247971-F 6-17Isotope/DPM LibrariesSetting Up A Quench Curve6in one rack and contain at least five and no

Page 24 - Using the Operating Controls

LS 6000 Scintillation System User’s ManualPN 247971-F 1-2 IntroductionGeneral DescriptionConfidence PackThe requires the Productivity Pack and include

Page 25 - 2.4 Using the Racks

LS 6000 Scintillation System User’s ManualPN 247971-F 6-18 Isotope/DPM LibrariesSetting Up A Quench Curve18. The standard data and a quench correlatio

Page 26 - Using the Racks

LS 6000 Scintillation System User’s ManualPN 247971-F 6-19Isotope/DPM LibrariesSetting Up A Quench Curve64. Highlight “Review/Edit/Delete Quench Curve

Page 27 - Setting Up the Racks

LS 6000 Scintillation System User’s ManualPN 247971-F 6-20 Isotope/DPM LibrariesSetting Up A Quench Curve6. To select the quench curve to edit, highli

Page 28

LS 6000 Scintillation System User’s ManualPN 247971-F 6-21Isotope/DPM LibrariesSetting Up A Quench Curve67. Highlight “Add or Delete Quench/Efficiency

Page 29 - 2.5 Preparing the Printer

LS 6000 Scintillation System User’s ManualPN 247971-F 6-22 Isotope/DPM LibrariesSetting Up A Quench Curveappropriate prompts and enter the values. To

Page 30 - 2.6 System Setup

LS 6000 Scintillation System User’s ManualPN 247971-F 6-23Isotope/DPM LibrariesSetting Up A Quench Curve61. From the Main Menu, highlight “Isotope/DPM

Page 31 - Energy Scale

LS 6000 Scintillation System User’s ManualPN 247971-F 6-24 Isotope/DPM LibrariesSetting Up A Quench Curve4. A prompt at the bottom of the menu allows

Page 32 - Allow Interrupt Count

LS 6000 Scintillation System User’s ManualPN 247971-F 7-1How Sample Preparation Affects ResultsChemiluminescence77How Sample Preparation Affects Resul

Page 33 - Figure 2.9 RS232 Setup Menu

LS 6000 Scintillation System User’s ManualPN 247971-F 7-2 How Sample Preparation Affects ResultsChemiluminescence5. Plant extracts containing chloroph

Page 34

LS 6000 Scintillation System User’s ManualPN 247971-F 7-3How Sample Preparation Affects ResultsRecognizing And Avoiding Statics7If you are unable to e

Page 35 - Auto DPM

LS 6000 Scintillation System User’s ManualPN 247971-F 1-3IntroductionGeneral Description1stepped into position for processing. In turn, each vial is r

Page 36 - 2.7 Calibration

LS 6000 Scintillation System User’s ManualPN 247971-F 7-4 How Sample Preparation Affects ResultsCounting Filters And PrecipitatesThis phase problem ca

Page 37

LS 6000 Scintillation System User’s ManualPN 247971-F 7-5How Sample Preparation Affects ResultsDistinguishing Sample And Instrument Variability7Proced

Page 38

LS 6000 Scintillation System User’s ManualPN 247971-F 7-6 How Sample Preparation Affects ResultsDistinguishing Sample And Instrument Variability1. The

Page 39 - Counting Samples

LS 6000 Scintillation System User’s ManualPN 247971-F 7-7How Sample Preparation Affects ResultsOther Factors That May Affect Accurate Results7Backgrou

Page 40 - Starting the Automatic Count

LS 6000 Scintillation System User’s ManualPN 247971-F 7-8 How Sample Preparation Affects ResultsOther Factors That May Affect Accurate Results

Page 41 - Conducting Count Single Rack

LS 6000 Scintillation System User’s ManualPN 247971-F a-1Instrument SpecificationsEfficiencyaaInstrument Specificationsa.1 EfficiencyThese specifica

Page 42

LS 6000 Scintillation System User’s ManualPN 247971-F a-2 Instrument SpecificationsPower Requirementsa.5 Power Requirementsa.6 Dimensionsa.7 Ambient T

Page 43

LS 6000 Scintillation System User’s ManualPN 247971-F a-3Instrument SpecificationsTemperature Control AccessoryaSet Point must be within 10oC of the

Page 44

LS 6000 Scintillation System User’s ManualPN 247971-F a-4 Instrument SpecificationsTemperature Control Accessory

Page 45

LS 6000 Scintillation System User’s ManualPN 247971-F b-1Installation RequirementsElectrical RequirementsbbInstallation RequirementsThis appendix de

Page 46 - 3.4 Auto DPM

LS 6000 Scintillation System User’s ManualPN 247971-F 1-4 IntroductionSpecificationsSystem SetupSystem Setup allows you to set system parameters tailo

Page 47 - 3.5 Multi-Task

LS 6000 Scintillation System User’s ManualPN 247971-F b-2 Installation RequirementsElectrical RequirementsPower lines within the laboratory should pro

Page 48 - Interrupt Count in Multi-Task

LS 6000 Scintillation System User’s ManualPN 247971-F b-3Installation RequirementsSpace Requirementsblines that are separate from those used to supp

Page 49

LS 6000 Scintillation System User’s ManualPN 247971-F b-4 Installation RequirementsEnvironment Requirements

Page 50 - Multi-Task

LS 6000 Scintillation System User’s ManualPN 247971-F c-1RS232SpecificationsccRS232c.1 SpecificationsStandard ASCII format, upper case only, is used

Page 51 - Interrupt Data

LS 6000 Scintillation System User’s ManualPN 247971-F c-2 RS232RS232 Data OutputWhen XOFF is received, data transmission stops until an XON is receive

Page 52

LS 6000 Scintillation System User’s ManualPN 247971-F c-3RS232RS232 Data Outputc<Length> is one printable ASCII character (Hex 20 through 7E o

Page 53

LS 6000 Scintillation System User’s ManualPN 247971-F c-4 RS232RS232 Data OutputHeader message includes a data item giving the type of file. There are

Page 54

LS 6000 Scintillation System User’s ManualPN 247971-F c-5RS232RS232 Input Modecc.4 RS232 Input ModeThe Input Mode of the LS functions as a remote ke

Page 55

LS 6000 Scintillation System User’s ManualPN 247971-F c-6 RS232RS232 Input Mode5 sec, Screen SelectionBS Back Space 1 secDL Delete 2 secPrinterPR Prin

Page 56

LS 6000 Scintillation System User’s ManualPN 247971-F c-7RS232RS232 Input ModecPercent, Window 1110 Requested Counting Precision, RealPercent, Windo

Page 57

LS 6000 Scintillation System User’s ManualPN 247971-F 1-5IntroductionPrecautions and Limitations1gloves can cause severe static problems which may pro

Page 58

LS 6000 Scintillation System User’s ManualPN 247971-F c-8 RS232RS232 Input Mode143 SPM: Number Of Data Points Integer144 SPM: Time Per Data Points (Mi

Page 59 - Setting Up User Programs

LS 6000 Scintillation System User’s ManualPN 247971-F c-9RS232RS232 Input Modec218 Units Isotope 2 Text (5 Char)219 Units Isotope 3 Text (5 Char)ID

Page 60

LS 6000 Scintillation System User’s ManualPN 247971-F c-10 RS232RS232 Input Mode314 Replicate Average CPM Coeff. of RealVariation, Percent, Isotope 13

Page 61 - 4.2 Editing A User Program

LS 6000 Scintillation System User’s ManualPN 247971-F c-11RS232RS232 Input ModecIsotope 3412 Average Reference Value, Isotope 1RealCoeff, of Variati

Page 62 - Editing A User Program

LS 6000 Scintillation System User’s ManualPN 247971-F c-12 RS232RS232 Input Mode513 DPM Ratio, Isotope 1 Over Isotope 2Real514 DPM Ratio, Isotope 3 Ov

Page 63

LS 6000 Scintillation System User’s ManualPN 247971-F c-13RS232RS232 Input Modec537 Xtal DPM Standard Average EfficiencyReal(%)MCA Spectrum DataData

Page 64

LS 6000 Scintillation System User’s ManualPN 247971-F c-14 RS232RS232 Input Mode905 Current System Status906 User Files (Data 0, .30, or 99 for all)In

Page 65 - Isotope 1

LS 6000 Scintillation System User’s ManualPN 247971-F c-15RS232RS232 Input Modecstopped, but counting will continue forthe remaining samples in this

Page 66 - Isotope 2 and Isotope 3

LS 6000 Scintillation System User’s ManualPN 247971-F c-16 RS232RS232 Input Mode170 Beta Norm Multiplier Isotope 1 Real171 Beta Norm Multiplier Isotop

Page 67 - Edit Other Parameters

LS 6000 Scintillation System User’s ManualPN 247971-F d-1Isotope Settings and Half-LifeddIsotope Settings and Half-LifeIsotope Setting and Half-Life

Page 68

LS 6000 Scintillation System User’s ManualPN 247971-F 1-6 IntroductionPrecautions and LimitationsMiniature VialsMiniature Vials within the dimensions

Page 69 - Background/Blank Subtraction

LS 6000 Scintillation System User’s ManualPN 247971-F d-2 Isotope Settings and Half-Life 113Sn-113mIn800 115.1 days-Sn; 99.48 minutes-In125I550 60.25

Page 70

LS 6000 Scintillation System User’s ManualPN 247971-F d-3Isotope Settings and Half-Lifed0,07 0,9526 0,66 0,6329 1,68 0,3121 4,30 0,05080,08 0,9461 0

Page 71 - Lum-Ex Correction

LS 6000 Scintillation System User’s ManualPN 247971-F d-4 Isotope Settings and Half-Life 0,36 0,7792 1,24 0,4234 3,10 0,1166 8,40 0,00300,37 0,7738 1,

Page 72 - Output Formats

LS 6000 Scintillation System User’s ManualPN 247971-F d-5Isotope Settings and Half-LifedFigure d.3 Conversion Factors.1 becquerel (Bq) therefore, =

Page 73

LS 6000 Scintillation System User’s ManualPN 247971-F d-6 Isotope Settings and Half-Life Figure d.5 Iodine-125 Half-Life: 60.25 daysDays0 2 4 6 8 10

Page 74

LS 6000 Scintillation System User’s ManualPN 247971-F d-7Isotope Settings and Half-Lifed140 0.3293 0.3242 0.3190 0.3140 0.3091 0.3043 0.2994 0.2947

Page 75 - Raw Data

LS 6000 Scintillation System User’s ManualPN 247971-F d-8 Isotope Settings and Half-Life Figure d.8 Carbon-14 Half-Life: 5730 yearsYears0 10 20 30 40

Page 76 - Collecting Spectral Data

LS 6000 Scintillation System User’s ManualPN 247971-F e-1Radioactive Material LicensingeeRadioactive Material LicensingThe following excerpts from t

Page 77 - 4.3 Copy User Program

LS 6000 Scintillation System User’s ManualPN 247971-F e-2 Radioactive Material Licensing(2) Shall assure that the device is tested for leakage of radi

Page 78 - Copy User Program

LS 6000 Scintillation System User’s ManualPN 247971-F e-3Radioactive Material Licensinge(‘1) Shall not export the device containing by-product mater

Page 79

LS 6000 Scintillation System User’s ManualPN 247971-F 1-7IntroductionHazards1Bio-VialsBio-Vials within the dimensions given on the left are compatible

Page 80 - 5.2 CPM/Xtal CPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-4 Radioactive Material LicensingRegional USNRC Office and Telephone Numberse.1 Regional USNRC

Page 81 - CPM/Xtal CPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-5Radioactive Material LicensingState Telephone & Address Numbersee.2 State Telephone &am

Page 82

LS 6000 Scintillation System User’s ManualPN 247971-F e-6 Radioactive Material LicensingState Telephone & Address NumbersConnecticut DirectorMonit

Page 83 - Results for CPM/Xtal CPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-7Radioactive Material LicensingState Telephone & Address NumberseIdaho Radiation Physici

Page 84

LS 6000 Scintillation System User’s ManualPN 247971-F e-8 Radioactive Material LicensingState Telephone & Address NumbersMaine Nuclear Engineering

Page 85 - 5.4 Auto DPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-9Radioactive Material LicensingState Telephone & Address NumberseMissouri ChiefBureau of

Page 86 - Results for Auto DPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-10 Radioactive Material LicensingState Telephone & Address NumbersNew Mexico ChiefBureau o

Page 87 - Setting Up DPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-11Radioactive Material LicensingState Telephone & Address NumberseDirectorBureau of Envi

Page 88

LS 6000 Scintillation System User’s ManualPN 247971-F e-12 Radioactive Material LicensingState Telephone & Address NumbersPennsylvania DirectorBur

Page 89 - 5.6 Xtal DPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-13Radioactive Material LicensingState Telephone & Address NumberseTexas DirectorIndustri

Page 90 - Xtal DPM

LS 6000 Scintillation System User’s ManualPN 247971-F 1-8 IntroductionHazards2. Contact your Radiation Health and Safety Officer for assistance. Furth

Page 91 - Setting Up Xtal DPM

LS 6000 Scintillation System User’s ManualPN 247971-F e-14 Radioactive Material LicensingState Telephone & Address Numbers‘ West Virginia ChiefRad

Page 92 - Loading Samples for Xtal DPM

LS 6000 Scintillation System User’s ManualPN 247971-F 2-1Getting StartedPowering Up22Getting Started2.1 Powering UpThe power switch for the LS is in a

Page 93 - 5.7 DPM % of Reference

Copyright, Licenses and Trademarks© Beckman Coulter, Inc., 2004. All rights reserved. No part of this publication may be reproduced, transcribed, tran

Page 94 - Setting Up DPM % of Reference

LS 6000 Scintillation System User’s ManualPN 247971-F 2-2 Getting StartedPowering UpThe only user accessible fuses are the power fuses, located next t

Page 95 - Xtal SL DPM % of Reference

LS 6000 Scintillation System User’s ManualPN 247971-F 2-3Getting StartedPower Failure2WARNING For continued protection against risk of fire, re-place

Page 96

LS 6000 Scintillation System User’s ManualPN 247971-F 2-4 Getting StartedOperating ControlsThe remaining keys on the keyboard are as follows:Operator

Page 97

LS 6000 Scintillation System User’s ManualPN 247971-F 2-5Getting StartedOperating Controls2DisplayA monitor is used to display the menus and instructi

Page 98 - 5.9 Single Photon Monitoring

LS 6000 Scintillation System User’s ManualPN 247971-F 2-6 Getting StartedOperating Controlsmenus. Below the Main Editing Window is the Data Entry Wind

Page 99

LS 6000 Scintillation System User’s ManualPN 247971-F 2-7Getting StartedUsing the Racks2Active Key Window. The monitor displays the count data and the

Page 100 - Single Photon Monitoring

LS 6000 Scintillation System User’s ManualPN 247971-F 2-8 Getting StartedUsing the Racksracks are supplied if Versa-Rack is installed. Bio-Vial racks

Page 101

LS 6000 Scintillation System User’s ManualPN 247971-F 2-9Getting StartedUsing the Racks2Setting Up the RacksIt is recommended that the Command Cards a

Page 102

LS 6000 Scintillation System User’s ManualPN 247971-F 2-10 Getting StartedUsing the RacksTo set up the Calibrate Rack, select one of the standard or m

Page 103 - Isotope/DPM Libraries

LS 6000 Scintillation System User’s ManualPN 247971-F 2-11Getting StartedPreparing the Printer2To set up the Sample Rack(s), select any of the supplie

Page 104 - Printing the Isotope Library

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FiSection 1 IntroductionIntended Use . . . . . . . . . . . . . . . . . . . . .

Page 105 - The Isotope Library

LS 6000 Scintillation System User’s ManualPN 247971-F 2-12 Getting StartedSystem SetupLoading the PaperRefer to the User’s Manual provided with the pr

Page 106 - Setup Menu

LS 6000 Scintillation System User’s ManualPN 247971-F 2-13Getting StartedSystem Setup2NOTE System Test is designed for use by a Beckman Coulter Autho

Page 107

LS 6000 Scintillation System User’s ManualPN 247971-F 2-14 Getting StartedSystem SetupPrinter PagingThe printer sets the top of page based on paper si

Page 108 - PN 247971-F

LS 6000 Scintillation System User’s ManualPN 247971-F 2-15Getting StartedSystem Setup2RS232The parameters for the RS232 port are selectable based on t

Page 109 - 5 minutes

LS 6000 Scintillation System User’s ManualPN 247971-F 2-16 Getting StartedSystem SetupTo change the settings, highlight “Stop Bits” and choose the app

Page 110 - 6.2 Setting Up A Quench Curve

LS 6000 Scintillation System User’s ManualPN 247971-F 2-17Getting StartedSystem Setup2Date/TimeThe current date and time is set at time of installatio

Page 111

LS 6000 Scintillation System User’s ManualPN 247971-F 2-18 Getting StartedCalibrationTo change these parameters, highlight “AUTO DPM Calibration Setup

Page 112 - Setting Up A Quench Curve

LS 6000 Scintillation System User’s ManualPN 247971-F 2-19Getting StartedTemperature Control Accessory2NOTE The unquenched 14C standard must be the f

Page 113

LS 6000 Scintillation System User’s ManualPN 247971-F 2-20 Getting StartedTemperature Control Accessory

Page 114 - Counting the Standards

LS 6000 Scintillation System User’s ManualPN 247971-F 3-1Counting SamplesPreparations for Use33Counting Samples3.1 Preparations for UseProvided the po

Page 115

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FiiStandard Racks (White) . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 116

LS 6000 Scintillation System User’s ManualPN 247971-F 3-2 Counting SamplesConducting An Automatic CountYou can load up to 28 standard vial racks or 36

Page 117

LS 6000 Scintillation System User’s ManualPN 247971-F 3-3Counting SamplesConducting Count Single Rack3NOTE If your samples have already been placed i

Page 118

LS 6000 Scintillation System User’s ManualPN 247971-F 3-4 Counting SamplesConducting Count Single RackCounting a Single Rack with the Default Paramete

Page 119

LS 6000 Scintillation System User’s ManualPN 247971-F 3-5Counting SamplesConducting Count Single Rack33. The Main Editing Window prompts you to load y

Page 120 - Editing Quench Curves

LS 6000 Scintillation System User’s ManualPN 247971-F 3-6 Counting SamplesConducting Count Single RackTo perform Count Single Rack using a User Progra

Page 121

LS 6000 Scintillation System User’s ManualPN 247971-F 3-7Counting SamplesConducting Count Single Rack31. With the Main Menu displayed, highlight “Coun

Page 122

LS 6000 Scintillation System User’s ManualPN 247971-F 3-8 Counting SamplesAuto DPMThe Main Editing Window then prompts you to load your samples and in

Page 123

LS 6000 Scintillation System User’s ManualPN 247971-F 3-9Counting SamplesMulti-Task3NOTE The Calibrate Rack must contain the unquenched l4~ standard

Page 124 - Deleting A Quench Curve

LS 6000 Scintillation System User’s ManualPN 247971-F 3-10 Counting SamplesMulti-TaskNOTE Multi-Task is automatically terminated if no user action (e

Page 125 - Manual Entry of Quench Curves

LS 6000 Scintillation System User’s ManualPN 247971-F 3-11Counting SamplesMulti-Task32. From the Multi-Task Menu, highlight “Interrupt Count” and pres

Page 126

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FiiiSection 3 Counting SamplesPreparations for Use . . . . . . . . . . . . . .

Page 127 - 7.1 Chemiluminescence

LS 6000 Scintillation System User’s ManualPN 247971-F 3-12 Counting SamplesMulti-TaskNOTE You may abort the Interrupt Count and resume Automatic Coun

Page 128 - Reducing Single Photon Events

LS 6000 Scintillation System User’s ManualPN 247971-F 3-13Counting SamplesMulti-Task3The User Program Review is displayed in the Supplementary Window

Page 129 - 7.3 Two Phase Samples

LS 6000 Scintillation System User’s ManualPN 247971-F 3-14 Counting SamplesMulti-Taskresults of the Interrupt Count are displayed and stored. Once sto

Page 130

LS 6000 Scintillation System User’s ManualPN 247971-F 3-15Counting SamplesMulti-Task3To access Interrupt Data:1. From either the Main Menu (Figure 3.1

Page 131 - Analysis of Data

LS 6000 Scintillation System User’s ManualPN 247971-F 3-16 Counting SamplesMulti-Task3. To print the data, highlight “Print Interrupt Data” and press

Page 132 - C Counting Efficiency

LS 6000 Scintillation System User’s ManualPN 247971-F 3-17Counting SamplesMulti-Task3Editing the Current User ProgramOnly the counting time or countin

Page 133 - Contamination

LS 6000 Scintillation System User’s ManualPN 247971-F 3-18 Counting SamplesMulti-Task7. Press PREVIOUS MENU to return to the Auto Count display.Using

Page 134

LS 6000 Scintillation System User’s ManualPN 247971-F 3-19Counting SamplesMulti-Task3Isotope Library Section 6 for adding and deleting isotopes if you

Page 135 - Instrument Specifications

LS 6000 Scintillation System User’s ManualPN 247971-F 3-20 Counting SamplesMulti-Task2. From the Multi-Task Menu, highlight “Isotope/DPM Libraries” an

Page 136

LS 6000 Scintillation System User’s ManualPN 247971-F 4-1Setting Up User ProgramsAbout the User Programs44Setting Up User Programs4.1 About the User P

Page 137

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FivSection 5 Data CalculationIntroduction . . . . . . . . . . . . . . . . . . .

Page 138 - Temperature Control Accessory

LS 6000 Scintillation System User’s ManualPN 247971-F 4-2 Setting Up User ProgramsAbout the User ProgramsFigure 4.2 Overview of User ProgramItem Defa

Page 139 - Installation Requirements

LS 6000 Scintillation System User’s ManualPN 247971-F 4-3Setting Up User ProgramsEditing A User Program4NOTE This section assumes you are familiar wi

Page 140 - 0.5 to 1.0 m in

LS 6000 Scintillation System User’s ManualPN 247971-F 4-4 Setting Up User ProgramsEditing A User Programthat selection, the allowable responses and th

Page 141

LS 6000 Scintillation System User’s ManualPN 247971-F 4-5Setting Up User ProgramsEditing A User Program4acceptable, use the Down Cursor key to move to

Page 142 - Environment Requirements

LS 6000 Scintillation System User’s ManualPN 247971-F 4-6 Setting Up User ProgramsEditing A User ProgramContinue onto the second Review/Edit Menu show

Page 143

LS 6000 Scintillation System User’s ManualPN 247971-F 4-7Setting Up User ProgramsEditing A User Program4CommentsThe Comments line provides descriptive

Page 144 - . Note that the output

LS 6000 Scintillation System User’s ManualPN 247971-F 4-8 Setting Up User ProgramsEditing A User ProgramChoose SPM (if installed) if you want to do si

Page 145

LS 6000 Scintillation System User’s ManualPN 247971-F 4-9Setting Up User ProgramsEditing A User Program4Edit Other ParametersIf no further editing is

Page 146 - Datum IDs

LS 6000 Scintillation System User’s ManualPN 247971-F 4-10 Setting Up User ProgramsEditing A User ProgramTo edit the other parameters, highlight “Edit

Page 147

LS 6000 Scintillation System User’s ManualPN 247971-F 4-11Setting Up User ProgramsEditing A User Program4Counting PrecisionCounting to a preset counti

Page 148

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FvLoading Samples for Xtal SL DPM% of Reference . . . . . . . . . . . . . . . .

Page 149

LS 6000 Scintillation System User’s ManualPN 247971-F 4-12 Setting Up User ProgramsEditing A User ProgramBlanks are samples that are counted at the be

Page 150

LS 6000 Scintillation System User’s ManualPN 247971-F 4-13Setting Up User ProgramsEditing A User Program4higher energy isotope into the lower energy i

Page 151

LS 6000 Scintillation System User’s ManualPN 247971-F 4-14 Setting Up User ProgramsEditing A User ProgramPhase MonitorThe 2 Phase Monitor detects and

Page 152

LS 6000 Scintillation System User’s ManualPN 247971-F 4-15Setting Up User ProgramsEditing A User Program4To change the settings, highlight “Output For

Page 153

LS 6000 Scintillation System User’s ManualPN 247971-F 4-16 Setting Up User ProgramsEditing A User ProgramFigure 4.10 Menu 2 for Edit Format: CPM.

Page 154

LS 6000 Scintillation System User’s ManualPN 247971-F 4-17Setting Up User ProgramsEditing A User Program4Figure 4.11 Menu 3 for Edit Format: CPM.Tabl

Page 155

LS 6000 Scintillation System User’s ManualPN 247971-F 4-18 Setting Up User ProgramsEditing A User ProgramTo select an item to include on the printout,

Page 156

LS 6000 Scintillation System User’s ManualPN 247971-F 4-19Setting Up User ProgramsCopy User Program4Either Sample or Compton Spectrum, neither spectru

Page 157

LS 6000 Scintillation System User’s ManualPN 247971-F 4-20 Setting Up User ProgramsCopy User ProgramProgram Section 4. The program copied is still sto

Page 158

LS 6000 Scintillation System User’s ManualPN 247971-F 5-1Data CalculationIntroduction55Data Calculation5.1 IntroductionThe Data Calculation programs a

Page 159

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-Fvi14C Counting Efficiency . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 160

LS 6000 Scintillation System User’s ManualPN 247971-F 5-2 Data CalculationCPM/Xtal CPMProcedures for setting up the data calculation programs are pres

Page 161

LS 6000 Scintillation System User’s ManualPN 247971-F 5-3Data CalculationCPM/Xtal CPM5Half-Life CorrectionSample data can be half-life corrected to th

Page 162

LS 6000 Scintillation System User’s ManualPN 247971-F 5-4 Data CalculationCPM/Xtal CPMCPM X Factor = Final Answer Printed OutTo change the factor for

Page 163

LS 6000 Scintillation System User’s ManualPN 247971-F 5-5Data CalculationSL CPM % of Reference/Xtal SL CPM % of Reference5If you are using replicates,

Page 164

LS 6000 Scintillation System User’s ManualPN 247971-F 5-6 Data CalculationSL CPM % of Reference/Xtal SL CPM % of ReferenceCPM/Xtal CPM Section 5 for i

Page 165

LS 6000 Scintillation System User’s ManualPN 247971-F 5-7Data CalculationAuto DPM5NOTE Do not leave an empty space for each missing replicate. One em

Page 166

LS 6000 Scintillation System User’s ManualPN 247971-F 5-8 Data CalculationDPMSetting Up Auto DPMWhen editing a User Program, from the Data Calculation

Page 167 - CHAPTER 1, PART 31

LS 6000 Scintillation System User’s ManualPN 247971-F 5-9Data CalculationDPM5Single Label DPM uses either IC# or H# Plus (if installed) as the quench

Page 168

LS 6000 Scintillation System User’s ManualPN 247971-F 5-10 Data CalculationDPMMAIN MENU and refer to Section 6.2 for setting up a quench curve for the

Page 169

LS 6000 Scintillation System User’s ManualPN 247971-F 5-11Data CalculationXtal DPM5Results for DPMTo count the samples, refer to Section 3.2. A typica

Page 170 - Region IV

Table of ContentsLS 6000 Scintillation System User’s ManualPN 247971-FviiState Telephone & Address Numbers . . . . . . . . . . . . . . . . . . . .

Page 171

LS 6000 Scintillation System User’s ManualPN 247971-F 5-12 Data CalculationXtal DPMFigure 5.9 Typical Printout for Dual Label DPM.Figure 5.10 Typica

Page 172

LS 6000 Scintillation System User’s ManualPN 247971-F 5-13Data CalculationXtal DPM5The Xtalscint DPM program assumes the samples are counting to the s

Page 173

LS 6000 Scintillation System User’s ManualPN 247971-F 5-14 Data CalculationXtal DPMStandard DateStandard Date is used to correct the dpm of the standa

Page 174

LS 6000 Scintillation System User’s ManualPN 247971-F 5-15Data CalculationDPM % of Reference5NOTE Do not leave an empty space for each missing replic

Page 175

LS 6000 Scintillation System User’s ManualPN 247971-F 5-16 Data CalculationXtal SL DPM % of ReferenceSingle and Dual Label DPM % of Reference (if inst

Page 176

LS 6000 Scintillation System User’s ManualPN 247971-F 5-17Data CalculationXtal SL DPM % of Reference5efficiency. This calculated efficiency is used to

Page 177

LS 6000 Scintillation System User’s ManualPN 247971-F 5-18 Data CalculationXtal SL DPM % of ReferenceXtal SL DPM % of Reference (if installed) is avai

Page 178

LS 6000 Scintillation System User’s ManualPN 247971-F 5-19Data CalculationXtal SL DPM % of Reference5If blanks are specified in the User Program as de

Page 179

LS 6000 Scintillation System User’s ManualPN 247971-F 5-20 Data CalculationSingle Photon MonitoringNOTE Do not leave an empty space for each missing

Page 180

LS 6000 Scintillation System User’s ManualPN 247971-F 5-21Data CalculationSingle Photon Monitoring5The Counting Time/Sample displayed is entered as th