LS 6000 Scintillation System User’s Manual
PN 247971-F
6-1
Isotope/DPM Libraries
The Isotope Library
6
6
Isotope/DPM Libraries
6.1 The Isotope Library
Introduction
The Isotope Library stores the half-life and window settings for the isotopes you wish to count.
These settings are accessed when the isotope(s) is specified during set up of a User Program.
The Isotope Library is divided into two parts. One section stores the settings for liquid
scintillators; the other section stores the settings for Xtalscint scintillators. Five common
radioisotopes for liquid and Xtalscint scintillators are permanently stored in the system. They
are
3
H,
14
C,
125
I,
35
S, and
32
P. Up to ten additional isotopes may be stored in the Isotope
Library under either section.
The window settings for a new isotope may be entered in either of two ways; the system can
count a sample of the isotope and record the settings automatically; or, if special settings for
the isotope are desired, they may be entered manually.
NOTE This section assumes you are familiar with the operating controls of the IS. If you are not familiar
with the operating controls, refer to Section 2.3.