Motorola LS6000B-U Guide de l'utilisateur Page 124

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 180
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 123
LS 6000 Scintillation System User’s Manual
PN 247971-F
6-22
Isotope/DPM Libraries
Setting Up A Quench Curve
appropriate prompts and enter the values. To display the previous menu, highlight the
first prompt and press the Up Cursor Arrow key.
12. Once all points are added or deleted and the background quench curve Is edited, press
START to print the new correlation table. Figure 6.13 shows a typical correlation table
for a SL
3
H quench curve.
13. Press PREVIOUSMENU] to store the edited quench curve. After storing the edited
curve, the menu shown in Figure 6.14 Is displayed.
If you do not want to store the changes after editing, press CANCEL. The changes are
not stored.
14. HighlIght “Review/Edit Complete” when you are finished editing and press SELECT.
Press MAIN MENU if you are finished using the DPM Library.
Deleting A Quench Curve
If a quench curve Is no longer required or if a new curve Is to be generated for a particular
Isotope, the existing curve may be deleted from the library. Deleting a quench curve affects all
User Programs using that quench curve.
To delete a quench curve:
Figure 6.17 Menu to Edit Background Quench Curve
Vue de la page 123
1 2 ... 119 120 121 122 123 124 125 126 127 128 129 ... 179 180

Commentaires sur ces manuels

Pas de commentaire